Download Inhoudsopgave Inhoud Print deze pagina
Inhoudsopgave

Advertenties

Beschikbare talen
  • NL

Beschikbare talen

  • DUTCH, pagina 97
Time course
1.
Press the Time Course menu option in the main menu.
2.
Press Options > More > Reading mode.
3.
Select Carousel 1 inch square or Carousel 10 mm square.
4.
In the toolbar, select the symbol for the carousel insert. This symbol
opens a window with carousel options; refer to
5.
Load the carousel insert according to the selected carousel option.
6.
The measurement interval between the individual measurements
must be at least 15 seconds.
7.
Start the measurement by pressing Zero or Read or Zero & Read
according to the selected carousel option.
Each zero solution and sample is measured in the defined
measurement interval and the curve is displayed.
Storing the time course
1.
Press Options > Select View and select the cell for which you
want to store the elapsed time.
2.
Select Table or Graph; the elapsed time is displayed.
3.
Press Options > Store Symbol and select one of the 20 cycle
storage locations.
4.
Press Store.
5.
The measurement data can be accessed in the main menu under
Recall Data > Time Course.
English 48

Accessories

Description
Carousel insert (sample changer) 10 mm
Carousel insert (sample changer) 1 inch
Table
2.
1-inch square cells, 8 matched cells, glass (10 mL)
10 mm square cells, matched pair, glass (3.5 mL)
10 mm square cells, quartz
10 mm disposable square micro cells (1.5 mL)
10 mm square plastic cells
Plug, neoprene for 1-inch square cells
Order no.
LZV902.99.00001
LZV902.99.00011
2495408
2095100
2624410
2629500
2743400
1480801

Advertenties

Inhoudsopgave
loading

Deze handleiding is ook geschikt voor:

Dr 6000

Inhoudsopgave