Nebulizer Kit with Advanced Valve Technology (AVT)

The C28P nebulizer is equipped with a breath-enhanced valve system
which allows the maximum amount of medication to be inhaled by the
patient. It optimizes the flow of medication during inhalation and reduces
medication loss during exhalation.
By synchronizing the flow of aerosol with the breathing, the therapeutic
effect of the inhaled medication is optimized.
Note: Only using the mouthpiece enables the synchronized functioning of
the nebulizer!
Inhalation phase
A
During inhalation, the inhalation valve (A)
D
opens and air enters the nebulizer.
C
The air is mixed with the generated
aerosol (B) and inhaled through the
mouthpiece by the patient (C).
In this phase the exhalation valve (D)
remains closed preventing aerosol from
being wasted.
B
EN
A
Exhalation phase
When exhaling through the mouthpiece,
D
the inhalation valve (A) of the nebulizer
closes, preventing aerosol to leave
the nebulizer. The aerosol generated
remains in the nebulizer and becomes
available for the next inhalation phase.
The exhaled air leaves the mouthpiece
through the exhalation valve (D).
5