9. Trouble Shooting
Q1 Power is not ON
Check if the plug is connected to mains.
Check for a blown fuse.
Q2 Low Pressure Alarm is on
Check if the Quick Connector is tightly secured.
Check if all tubing connections along mattress are secured.
Check if the air hoses are kinked or broken.
Check if the CPR knob is tightly closed.
Q3 Power Failure Alarm is on
Check if the power is suddenly shut down.
Check if the power cord is connected properly.
Q4 Patient is bottoming out
Pressure setting might be inadequate for the patient, increase comfort range 1 level
higher
and wait for a few minutes for best comfort.
Q5 Mattress form is loose
Check if all the snap buttons or straps of mattress are all securely fastened.
Check if the mattress is fixed to the bed frame by straps.
Q6 No air released from some air outlets of the air tube connector
This is in alternating mode. The air is released through outlets by turns at each cycle.
If the problems you have are not solved by the above information, please contactlocal
representative or place of purchase.
10. Error Code Table
Code No.
00
01-06
08-12
16
17
18
19
Initial inflation overtime
Low pressure
System malfunction
Pressure control system malfunction
System access control malfunction
Mattress disconnected
Connect to the wrong mattress
Description