Commissioning

Switch the washer-disinfector on using the  button.
Press the push button on the module.
If the control LED lights up red, the module has been correctly in-
stalled and you can activate the serial interface and start the configur-
ation.
If the control LED does not light up:
- The module has not been correctly installed.
Press the module gently to check whether it has connected cor-
rectly
- The washer-disinfector is not switched on.
Switch the washer-disinfector on using the  button.
- There may be a fault with the module
Contact Miele Service.

Configuring the interface

Please observe the notes in the instruction manual for your washer-
disinfector under "Further settings/Interface".
en - Commissioning