USE
General
Make all vacuum manifold and electrical connections and
refer to Turbo-V ICE pump instruction manual before to
operating the Turbo-V ICE contr oller.
!
To avoid injury to personnel and damage to the
equipment, if the pump is laying on a table make sure it
is steady. Never operate the Turbo-V ICE pump if the
pump inlet is not connected to the system or blanked
off.
CAUTION!
The pump is factory set with Argon process gas. To
avoid pump failures, do not select Air if the pump is
used with heavy gases.
NOTE
The input signal P1 connector should be left in position
including the shipping links if no external connections
are
made.
The
Turbo-V ICE pump can be switched on at the same
time.
NOTE
When the Turbo-V 2000 ICE pump is baked by a
membrane pump, the Soft Start mode should be
deselected.
Startup
Plug the controller power cable into a suitable power
source.
The display lights, and shows:
A U T O T E S T
C H E C K
NOTE
The above message is only displayed when a board
with optional serial ports is installed.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
WARNING!
forepump
and
the
A U T O T E S T
R E A D Y
S O F T
S T A R T
The controller with the Soft Start mode allows the pump to
ramp-up to Normal Speed slowly with a minimum ramp-up
time of 80 minutes and a maximum of about 110 minutes.
The Soft Start mode is always operative as default mode. If
it is necessary to deselect this mode refer to the following
paragraph.
If the Soft Start mode is deselected, the ramp- up will be
done within 10 minutes and the display changes as
follows:
P U M P
R E A D Y
S T A R T
If the pump is not connected, the display will be as shown :
C H E C K
T O
After pressing the START push-button, if the P1
connector is not in place with the link or the external
interlock connection are open, the display will be as
shown in the following figure.
P U M P
W A I
I
N T E R L O C K
Unplug the controller power cable and verify the P1
connection according to paragraph "Interconnections".
80
TECHNICAL INFORMATION
O K
F O R
L O C A L
:
P U S H
B U T T O N
NOTE
C O N N E C T
I O N
P U M P
NOTE
T
I
N G
87-900-929-01 (H)
H S
H S
H S