Download Inhoudsopgave Inhoud Print deze pagina

Fuse Heater Jacket; Interconnections; Connection J1 - Logic Input/Output Interconnections - Varian Turbo-V 2000 ICE Bedrijfshandleiding

Inhoudsopgave

Advertenties

Beschikbare talen
  • NL

Beschikbare talen

  • DUTCH, pagina 28

FUSE HEATER JACKET

The following figure shows the location of this
assembly.
Rear panel

INTERCONNECTIONS

The following figure shows the Controller inter-
connections.
Controller models 969-9448
and 969-9449 interconnection
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Connection J1
Logic Input/Output Interconnections
FEMALE CO NNE C T O R
13
25
12
24
11
23
10
22
9
21
8
20
7
J1
19
6
I/O
18
5
17
4
16
3
15
2
14
1
OPEN COLLECTOR OUTPUTS IMAX = 60mA
POWER MONITOR : 10V = 1000W
S PEED M O NIT O R : 1 0 V = 33 KRPM
J1 input connector
All the logic input to the controller must be connected at
J1 mating connector. With the provided J1 mating
connector (shipped with pin 3 and pin 8 shorted) make
the connections with AWG 24, (0.24 mm
wire to the pins indicated in the figure to obtain the
desired capability. The following table describes the
signals available on the connector.
PIN
1-14 Remote START/STOP S1 optically isolated from
the internal circuit, requires a permanently closed
contact (relay contact. transistor, etc.). When the
contact closes the turbopump starts at high speed
and when the contact opens, the turbopump is
stopped. With the remote mode operation
selected,
the
inoperative.
2-15 Remote START/STOP S2 optically isolated from
the internal circuit, requires a permanently closed
contact (relay contact. transistor, etc.). When the
contact closes the turbopump starts at low speed
and when the contact opens, the turbopump is
stopped. With the remote mode operation
selected,
the
inoperative.
3-16 INTERLOCK optically isolated from the internal
circuit, this signal can be used to delay the starting
of the turbopump. It requires a permanent closed
contact before starting the turbopump.
4-17 Remote/Local signal 24 V, 60 mA, optically
isolated output (pin 17 positive, pin 4 negative).
The output Voltage will be present when the
controller is configured in Remote Mode.
5-17
START S1 signal 24 V, 60 mA, optically isolated
output (pin 17 positive, pin 5 negative). The output
Voltage will be present when the START push-
button on front panel is pressed or the remote start
is present. or the function has been requested by
RS 232/422/485, until NORMAL operation is
reached.
72
TECHNICAL INFORMATION
SPEED MONITOR (0 TO 10V)
GND
POWER MONITOR (0 TO 10V)
BRAKING (OPEN COLLECTOR)
FAILURE (OPEN COLLECTOR)
+24VDC
OVERTEMP (OPEN COLLECTOR)
+24VDC
PURGE FLOW (OPEN COLLECTOR)
+24VDC
NORMAL S2 (OPEN COLLECTOR)
+24VDC
NORMAL S1 (OPEN COLLECTOR)
+24VDC
STARTING S2 (OPEN COLLECTOR)
+24VDC
STARTING S1 (OPEN COLLECTOR)
+24VDC
REMOTE/LOCAL (OPEN COLLECTOR)
GND
INTERLOCK
GND
START S2
GND
START S1
2
) or smaller
DESCRIPTION
front
panel
push-button
front
panel
push-button
87-900-929-01 (H)
is
is

Advertenties

Inhoudsopgave
loading

Gerelateerde Producten voor Varian Turbo-V 2000 ICE

Deze handleiding is ook geschikt voor:

969-9448969-9449

Inhoudsopgave