Zeon Endeavour MD
Weight of battery pack
7.06-7.13 ounces (200-202 grams)
Dimensions of battery pack
Height = 3.6 inches (92 mm) | Width = 2.1 inches (55 mm) | Depth = 1.1 inches (28 mm)
Battery type
Li-Polymer, 3.6-3.63 VDC (6500-6600 mAh)
Light intensity modes
Level 1 = 53 lumens | Level 2 = 68 lumens | Level 3 = 85 lumens
Battery life
Level 1 = 24 hours | Level 2 = 17 hours | Level 3 = 10 hours
Charge time (each battery pack)
6 hours
Color Rendering Index (CRI)
CRI 90+
Correlated Color Temperature (CCT)
CCT 5700K
Weight of the headlight
.25 ounces (7 grams)
Diameter of headlight
.55 inch (14 mm)
Length of headlight cable
56 inch (142 cm)
Electrical
Class II
Operation mode
Continuous
Standards
EMC = IEC60601-1-2, class B, EN55011 | Safety = IEC60601-1, IEC 62471
Operating environment
32° to 77°F (0° to 25°C) | 10% to 85% relative humidity (non-condensing) | 0.5 atm to 1.0 atm (500 hPa to 1060 hPa) atmospheric pressure
CISPR 11 – Emissions: Radiated RF
IEC 61000-3-2 – Emissions: Harmonic
IEC 61000-3-3 – Emissions: Voltage Fluctuations / Flicker
IEC 61000-4-2 – Immunity: Electrostatic Discharge
IEC 61000-4-3 – Immunity: Radiated RF
IEC 61000-4-3 – Immunity: Proximity Fields from RF Wireless Communications Equipment IEC 60601-1-2:2014, section 8.10
The EMISSIONS characteristics of this equipment make it suitable for use in industrial areas and hospitals (CISPR 11 class A). If it is used in a residential environment (for which CISPR 11 class B is normally required) this equipment might not offer adequate
protection to radio-frequency communication services. The user might need to take mitigation measures, such as relocating or re-orienting the equipment.
INSTRUCTIONS FOR USE AND CARE
Orascoptic ™ Zeon ™ Endeavour ™ MD Portable LED Headlight
Specifications
Zeon Endeavour MD (Generation 2)
Level 1 = 68 lumens | Level 2 = 90 lumens | Level 3 = 115 lumens
CRI 92+
CCT 6500K
Electromagnetic Compatibility Test
Group 1, Class A
IEC 61000-4-4 – Immunity: Electrical Fast Transient / Burst
Class A
IEC 61000-4-5 – Immunity: Surge
Complies
IEC 61000-4-6 – Immunity: Conducted RF
±8 kV contact | ±15 kV air
IEC 61000-4-8 – Immunity: Power Frequency Magnetic Fields
3 V/m, 80 MHz – 2.7 GHz, 80 % 1 kHz AM
IEC 61000-4-11 – Immunity: Voltage Dips and Short Interruptions
See back cover for Customer Care or Technical support contact information | Customers outside our direct sales regions, contact your local distributor
±2 kV for power supply lines | ±1 kV for input/output lines
±1 kV differential | ±2 kV common
3 Vrms, 150 kHz – 80 MHz, 80 % 1 kHz AM
6 Vrms, ISM and Amateur Radio bands between 150 kHz – 80 MHz, 80 % 1 kHz AM
30 A/m, 50/60 Hz
<5 % UT for 0.5 cycles | 40 % UT for 5 cycles | 70 % UT for 25 cycles | <5 % UT for 5 seconds