Each measurement is automatically stored in the memory. The memory
has a capacity for 90 measurements. When the memory is full, each
new measurement will erase the oldest stored measurement.
Checking the memory:
Press the R button to display the average of the last 3 measure-
ments.
Press the R button again, '01' will appear followed by the results
of the last measurement.
Press the R button again, '02' will appear followed by the results
of the second-to-last measurement.
Repeat this for all previous measurements.
Erasing the memory:
Make sure to turn on the FB-180.
Press and hold the R button for approx. 5 seconds until 'CLr'
appears in the display. All measurements stored in the memory
will now be deleted.

POSSIBLE PROBLEMS AND SOLUTIONSGELIJKE

No reaction when turning on the FB-180:
Batteries empy? Install fresh ones.
"Err" appears in the display:
Impossible to pressurize the cuff. Check whether it's loose and
check the connection between the cuff and blood pressure mon-
itor.
The air escapes too quickly. If this happens repeatedly, please
contact the service department of Fysic.
Movement detected during the measurement. Please don't move
during the measurement.
Interfering signals near the FB-180. Remove any sources of inter-
ference, such as mobile phones.
Incorrect measurement value. The cuff is incorrectly attached.
Please attach the cuff again.
MEMORY