EMC RICHTLIJNEN
Tabel 2
Guidance and manufacturer's declaration – electromagnetic Immunity
IEC 60601-1-2
Immunity Test
Test level
Electrostatic
±8 kV contact
discharge (ESD)
±2 kV, ±4 kV, ±8 kV, ±15 kV air
IEC 61000-4-2
Electrical fast
±2 kV for power supply lines
±1 kV signal input/output
transient/burst
100 kHz repetition frequency
IEC 61000-4-4
±0.5 kV, ±1 kV differential mode
Surge
±0.5 kV, ±1 kV, ±2 kV common mode
IEC61000-4-5
Voltage dips, short
0% U
; 0,5 cycle. At 0°, 45°, 90°, 135°,
T
interruptions and
180°, 225°, 270° and 315°.
voltage variations
0% U
; 1 cycle and 70% U
T
on power supply
25/30 cycles; Single phase: at 0°.
input lines
0% U
; 250 / 300 cycle
T
IEC 61000-4-11
Power frequency
30 A/m
magnetic field
50 Hz / 60 Hz
IEC 61000-4-8
3 V
Conduced RF
0,15 MHz – 80 MHz
IEC61000-4-6
6 V in ISM and amateur radio bands
between 0,15 MHz and 80 MHz
80% AM at 1 kHz
10 V/m
Radiated RF
80 MHz – 2,7 GHz
IEC61000-4-3
80% AM at 1 kHz
NOTE U
is the a.c. mains voltage prior to application of the test level.
T
30
Compliance level
±8 kV contact
±2 kV, ±4 kV, ±8 kV, ±15 kV air
±2 kV for power supply lines
Not Applicable
100 kHz repetition frequency
±0.5 kV, ±1 kV differential mode
Not Applicable
0% U
; 0,5 cycle. At 0°, 45°, 90°, 135°,
T
180°, 225°, 270° and 315°.
;
0% U
; 1 cycle and 70% U
;
T
T
T
25/30 cycles; Single phase: at 0°.
0% U
; 250 / 300 cycle
T
30 A/m
50 Hz / 60 Hz
3 V
0,15 MHz – 80 MHz
6 V in ISM and amateur radio bands
between 0,15 MHz and 80 MHz
80% AM at 1 kHz
10 V/m
80 MHz – 2,7 GHz
80% AM at 1 kHz
Tabel 3
Guidance and manufacturer's declaration - electromagnetic Immunity
Radiated RF
Test
Band
Service
Frequency
(MHz)
IEC61000-4-3
(MHz)
(Test
specifications
385
380-390
for
TETRA
ENCLOSURE
400
PORT
IMMUNITY to
450
430-470
GMRS 460,
RF wireless
FRS 460
communicati-
ons
710
704-787
LTE Band
equipment)
13,
745
17
780
GSM
810
800-960
800/900,
870
TETRA 800,
iDEN 820,
930
CDMA 850,
LTE Band 5
GSM 1800;
1720
1700-
1990
CDMA 1900;
GSM 1900;
1845
DECT;
LTE Band 1,
1970
3, 4,25;
UMTS
Bluetooth,
2400-
2450
WLAN,
2570
802.11
b/g/n, RFID
2450, LTE
Band 7
5100-
5240
WLAN
5800
802.11
5500
a/n
5785
EMC RICHTLIJNEN
IEC
Modulation
Distance
Maximum
60601-1-2
(m)
Power
Test Level
(W)
(V/m)
Pulse
1.8
0.3
27
modulation
18 Hz
FM ± 5k Hz
2
0.3
28
deviation
1 kHz sine
Pulse
0.2
0.3
9
modulation
217 Hz
Pulse
2
0.3
28
modulation
18 Hz
Pulse
2
0.3
28
modulation
217 Hz
Pulse
2
0.3
28
modulation
217 Hz
Pulse
0.2
0.3
9
modulation
217 Hz
Compliance
level
(V/m)
27
28
9
28
28
28
9
31