ibp_TD4222_GB
07.01.2005
Model
Measuring unit
Sample material
Sample volume
Measuring range
Measuring time
Measured results
Display
Memory
Operating temperature
Storage/transport temperature
Power supply
Battery operating life
Dimensions
Weight
CE certification
We reserve the right to make changes to technical details without notification for reasons of updating.
All manuals and user guides at all-guides.com
8:03 Uhr
Seite 33
:
ibp TD 4222
:
switchable between mg/dL and mmol/L
:
capillary whole blood
:
2 microlitres
:
20 to 600 mg/dL (1.1 to 33.3 mmol/L)
:
10 seconds
:
whole blood-calibrated
:
large liquid crystal display (LCD)
:
450 measured results with time and date
:
+10 °C to +40 °C,
10 to 90 % relative humidity (non-condensing)
:
-20 °C to +70 °C, 5 to 95 % relative humidity
:
1 x 3V CR2032 lithium battery
:
for around 1000 measurements
:
78 x 46 x 17 mm
:
approx. 40 g (including batteries)
:
The device is certified to 98/79/EC, for "in vitro" diagnostic
devices and the following standards: IEC 60601-1,
IEC 61010-1, IEC 61010-1-2, IEC 61326 and ISO 15197
specifications 15
33