Speed and RCF
= Toggle between RPM
and RCF; activate
speed setting at
increment of 10
Acceleration and
deceleration
= Toggle between
acceleration and
deceleration; activate
Accel/Decel setting
(level 0-9)
Precooling
= Precool the centrifuge
and rotor to the preset
temperature (cooled
models only)
Prog
= Recall stored
programs
Store
= Store settings as
programs including the
rotor used (99
programs)
Quick
= For short runs
2.3 General operations
Notes:
⚫ Placed the To re-start the centrifuge after a run, one must open and close the lid
once (If this is not the case, the word "open" inside the rpm/rcf-display will flash),
which is a build-in safety feature.
⚫ All liquid handling must be performed outside of the centrifuge to avoid spilling.
⚫ Before using the emergency lock release, please make sure the rotor has stopped
completely by check through the see-through port on the lid.
⚫ Use emergency lock release to open the lid as described in the manual if the
safety lock was triggered.
⚫ All places of swing out rotors must be occupied with at least buckets and racks.
5
Temperature
= Activate temperature
setting from -20°C to
40°C (cooled models
only)
Time
= Activate time setting
(max. 99h and 59min, or
continuous)
Start
= Start the
centrifugation (only start
if the lid is securely
closed)
Stop
= The unit can be
stopped at any time
Door Open
= Open centrifuge lid
(only open if the rotor
has stopped completely)
Dialing knob/dial
=Dialing knob for
adjusting the activated
values