Cleaning
The walker can be cleaned with mild household detergents. Running liquids should be avoided. Do not use scrub
detergents, steam- or high-pressure cleaners. The walker cannot be hosed because of the electric motor.
In case of multiple patients using the walker, it is necessary to clean and disinfect the walker and its accessories be-
fore use by another patient. Remove the battery from the walker before the disinfection/cleaning process begins.
Allow the walker to dry in room temperature before use. Do not store below 0°C before walker is completely dry.
Cleaning is done manually in use areas without mechanical units (e.g., ultrasonic cleaners or washer-disinfectors).
No further use of products in the case of water penetration.
Disinfection
Disinfection by wiping generally accessible surfaces with disinfectant. Disinfection must only be carried out by
authorized personnel and by wearing sufficient protective equipment. The surface must be cleaned with at disinfec-
tion that contains 70–80 % ethanol. We advise against disinfection containing chlorine and phenol. Running liquids
should be avoided. The battery of the walker must be removed before the disinfection process take place. Before
use, allow the Walker to air-dry in room temperature until there is no more superficial moisture on the product. The
manufacturer cannot be held responsible for any damage or wounds that might occur by the use of harmful disinfection
or disinfection carried out by unauthorized personnel.
Inspection // Maintenance // Re-use
Do not service and maintain this Medical Electrical Equipment while in use!
The wheel bearings are sealed and maintenance-free. The other parts do not need to be lubricated. It is rec-
ommended that general inspection and maintenance is carried out and that the items below are checked on a
regular basis, and at least once a year (how frequent is determined by how and how often the walker is used):
Tubes, frame and screws, cables, handles and forearm supports, brakes, wheels and accessories. This also ap-
plies when the walker is to be prepared for re-use. Please see detailed instructions in this user manual about
maintenance of the brakes and cleaning and disinfection of the walker.
If the product does not work properly
1 = Possible reason
Brakes
1 Worn out / defect wheel
1 Not braking effectively
1 Worn out / defect brake mechanism on wheel
1 Worn out / defect brake handle
1 Worn out / damaged brake wire
Height adjustment
1 Battery is flat/does not work
1 The height adjustment does not work
It only goes down, not up
1 The battery is not correctly mounted
1 Loose cable contact
1 Damage on the cable
1 Defect switch
1 Defect height adjusting panel (accessory)
In the event that all of the above has been checked and the product is still not operating correctly, then the
cause may be damage to the control box or actuator. Please contact your supplier.
16
TAURUS E
2 = Action
2 Change the complete wheel
2 Adjust the brake wire
2 Change the entire back wheel section
2 Change the complete brake handle
2 Change the brake wire
2 Charge or change the battery.
See point 9; charging the battery
2 See point 9; charging the battery
2 Check that the battery is clean and correctly mounted
2 Check that all contacts are properly in place and
tightly fitted
2 Check all cables and replace any damaged one(s)
2 Change the switch
2 Change the height adjusting panel
www.topromobility.com