9 Technical Specifications

Model: B26 / FB150
Weight: 261.3g (Batteries included)
Display: 48*65mm 2.56"x1.89" LCD Digital Display
Size: 118 (W) x 110 (L) x 52 (H) mm 4.65"(W)x4.33"(L)x2.05"(H)
Accessories: 1×Main Device, 1×Cuff, 1×Users manual
Operating Conditions: Temperature: 5°C to 40°C; Humidity: 15% to 93% RH;
Storage And Shipping Conditions: Temperature: -25°C to 70°C; Humidity:
Atmospheric pressure range 70kPa~106kPa
Measuring method: Oscillometric Pressure sensor: Resistive
Measuring range: DIA: 40-130mmHg; SYS: 60-230mmHg Pulse: 40 to 199 per minute
Cuff pressure display range: <300mmHg
Memory: Automatically stores the last 120 measurements for 2 users (total 240)
Measuring resolution: 1 mmHg
Accuracy: Pressure within ± 3 mmHg / pulse ± 5 % of the reading Power source: a) 4*AAA batteries, 1.5
V
b) AC adapter INPUT 100-240VAC 50/60HZ OUTPUT 5V DC 1A Accessories: Wide range rigid cuff
8.7" – 16.5" (22 - 42 cm)
Automatically power off: 60 seconds
Users: Adult

10 EMC Declaration

1) *This product needs special precautions regarding EMC and needs to be installed and put into
service according to the EMC information provided, and this unit can be affected by portable and
mobile RF communications equipment.
2) * Do not use a mobile phone or other devices that emit electromagnetic fields, near the unit. This
may result in incorrect operation of the unit.
3) * Caution: This unit has been thoroughly tested and inspected to assure proper performance and
operation!
4) * Caution: this machine should not be used adjacent to or stacked with other equipment and that if
adjacent or stacked use is necessary, this machine should be observed to verify normal operation in
the configuration in which it will be used
Guidance and manufacture's declaration – electromagnetic immunity
The device is intended for use in the electromagnetic environment specified below. The customer
or the user of device should assure that it is used in such an environment.
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
IEC 60601 test
level
±8 kV contact ±2
kV, ±4 kV, ±8 kV,
±15 kV air
Compliance
level
±8 kV contact ±2
kV, ±4 kV, ±8 kV,
±15 kV air
15
Electromagnetic environment -
guidance
Floors should be wood, concrete
or ceramic tile. If floor are
covered with synthetic material,
the relative humidity should be at

Advertenties