Cautions:
Never sterilize the Adapter Cable by any means.
You can immerse the patient end of the Adapter Cable, but no longer than specified by the cleaning
agent or disinfectant manufacturer. Do not immerse the Fetal Monitor connector.
Select recommended cleaning agents and disinfectants carefully, because some products with similar
names have different compositions.
Always dilute cleaning agents and disinfectants according to manufacturers' instructions.
Do not use strong oxidants or bleaches containing sodium hypochlorite (e.g. Clorox™).
Do not use Povodine
Do not use disinfectants containing iodine complexes.

Exterior Cleaning and Disinfection

Perform the following procedure after every use:
Step
Cleaning and Disinfecting After Every Use
Press the cover over the Flush-Cleaning Port
1
assure it is closed.
2
Spray cleaning agent directly onto the Adapter Cable, or
onto a soft, clean cloth. You can also immerse the patient
end of the Adapter Cable.
3
Wipe all Adapter Cable surfaces clean from one end to
the other.
4
Change to a dry cloth and dry the Adapter Cable com-
pletely.
5
Repeat steps 1-4 using a recommended disinfectant.
6
Using a clean, dry cloth, wipe all residual disinfectant
from the Adapter Cable.
WARNING: Any residual disinfectant left on the cable
could cause severe skin irritation or blistering if it
comes in contact with a patient.
7
If you immersed the patient end of the Adapter Cable,
rinse the internal cavity following
procedure, Cleaning the Internal Cavity.
8
Check for signs of any deterioration or damage. If dam-
age is noticed, discard the Adapter Cable immediately.
®
®
, Sagrotan
, Mucocit
steps 4 and 5 in the next
®
-P or strong organic solvents.
to
Flush-
Cleaning
Port
4
Patient End of
Adapter Cable
Spiral Electrode
Input
Snap-Connector
(not visible)
Cable-Connector
to Fetal Monitor