Congratulations on your purchase of the Aqua Inspektor
electronic pool tester. The tester is a battery-operated mea-
suring device which determines the specific dissolved matter
together with the included reagent tablets. The tester and
the tablets are only intended for the designations
mentioned in the table in non-commercial applications and
may not be used for other purposes.
Please read through this manual and the special instructions
carefully before using it for the first time.
Important notes
Tablet reagents are intended for chemical water testing and
analysis only and may not be used for other purposes.
Caution! Potentially harmful if the tablet reagents are
swallowed.
Do not swallow! Keep the tablet reagents out
of the reach of children.
Before you start
Insertion and replacement of batteries:
Battery type 2 x AAA, LR 03, 1.5 V.
Only perform when the device is dry: Open the battery
compartment on the underside of the device. Pay attention
to the correct polarity of the batteries. Make sure that the
battery compartment cover is firmly seated when closed.
Replace the batteries if this symbol appears in the
display.

Error messages

No. Meaning and possible causes
01
Insufficient light at ZERO. Poss. cause: Duct contami-
nated or light path blocked; sample strongly coloured?
03
Division by 0. Poss. cause: Duct contaminated or light
path blocked; sample strongly coloured?
20
Too much light at detector. Poss. cause:
Cap not applied; measurement in bright sunlight
21
Too much light at detector. Poss. cause:
Cap not applied; measurement in bright sunlight
22
LowBat occurred during measurement.
Change battery, clean contacts
• Maintenance
1. Keep the instrument in a clean condition.
2. No solvent, abrasive materials or brushes should be used
to clean the instrument. Clean the sample chamber using
a soft tissue.
3. Dry the instrument when not in use, store under ambient
environmental conditions.
4. Remove battery if instrument is being stored for an
extended period.
• Trouble shooting
1. Use only blister tablets with black printing.
2. Do not touch the tablets. This can cause contamination
and give pour results.
3. Close the blue lid on top of the sample chamber tightly
before zeroing/reading.