Add-on module E 702/1
Items supplied:
Add-on module E 702/1 for E 701/2 and E 525/2 (H 160, W 530, D 560 mm)
,
Please read the Warning and Safety instructions in the Operating
Instructions for your washer-disinfector carefully. You should only process
instruments/medical products in this unit if they have been declared as suitable
for machine processing by the manufacturer. Please also observe the
manufacturer's care instructions. The programme selected must be suitable for
the materials from which the medical products to be processed are made.
Where possible the programme should include thermal disinfection. The final
rinse should, if possible, always be carried out with fully demineralised water.
If using automatic mobile unit recognition, please make sure that the coding is
correctly set.
Areas of application:
Additional mesh trays containing surgical instruments can be processed in mobile
units 701/2 and E 525/2 when the add-on module is attached.
This requires the add-on module to be connected to the water intake for the mobi-
le unit.
,
Please also observe the mobile unit Operating instructions for surgical
instruments.
Before loading the machine and before starting a programme, please check:
– that the spray arms rotate freely,
– that the spray arm jets are clean and free of any blockages,
– that the mobile unit is correctly fitted into the cabinet and that it has connected
correctly with the water intake.
To fit the add-on module
^ Unscrew nuts a on the securing screws on the mobile unit with an SW 10
spanner, and remove the spring washers.
^ Remove blanking plate b from the water intake.
Keep blanking plate b, nuts a and the spring washers in case the add-on
module is removed at a later date.
^ To connect the add-on module, first unscrew nuts c from the fixing screws on
add-on module d with an SW 10 spanner, and remove the spring washers.
^ Place add-on module d on top of the mobile unit. See illustration above for
installation details.
After replacing the fixing screws:
^ Replace the spring washers on the fixing screws, and then nuts c, and tighten
with an SW 10 spanner.
Miele Company Ltd., Fairacres, Marcham Road, Abingdon, Oxon, OX14 1TW, Tel 0330 160 6600,
Internet: http://www.miele.co.uk / ! 0297
Alteration rights reserved / Publication date 2015-11-01
E 701/2 fitted with E 702/1 add-on module
For validation purposes please make sure that you follow the loading
instructions given on the template.
To tighten the fixing screws
A:
^ Slot the side hooks on the add-on module over the outer frame of the mobile unit
for surgical instruments.
^ Lower the add-on module onto the mobile unit, and secure to the mobile unit
through the drill holes using the fixing screws. See B in illustration above.
B:
^ Tighten fixing screws with an SW 7 spanner.
en