Download Inhoudsopgave Inhoud Print deze pagina

Commissioning; Unit Installation; Connecting And Activating The Device; Temperature Setting - VOLTCRAFT TC-150 Gebruiksaanwijzing

Temperatuurkalibrator
Inhoudsopgave

Advertenties

Beschikbare talen

Beschikbare talen

COMMISSIONING

Unit Installation

Place the device on a level, heat-resilient surface. The storage area may heat up slightly
when air escapes. Keep fl ammable objects away from the calibration body (5). Make sure
that there is suffi cient air circulation. A minimum distance of 20 cm must be complied with
on all sides.

Connecting and Activating the Device

Plug the included mains cable into the mains connection (14) at the back of the device and into a mains
socket with a protective ground.
Switch on the device with the operating switch (15).
Switch position "ON" (I)
Switch position "OFF" (0)
The device fan runs and a system test is performed. After about 3s, the last set target and actual
temperatures are displayed.

TEMPERATURE SETTING

The cooling and heating surfaces at the front heat up in operation. There is danger
of burning at touch!
The desired temperature can be set at the temperature calibrator with two buttons.
Button "UP" (7) increases the temperature target value. Button "DOWN" (8) reduces the target value.
Every push of a button switches the temperature up or down by 0.1 °C. Long pushing of one of these
buttons permits fast adjustment. Adjustment takes pace in three speed levels, depending on how long
the button is pushed.
During setting, the green target value display (2) fl ashes quickly (fl ickering). This shows that you are in
setting mode. The temperature is not adjusted yet.
When you have set the desired temperature, confi rm your input with the "SET" button (10). The green
target value display (2) continues to be lit.
The red actual value display slowly adjusts to the target value.
The set area is larger than the control area for system reasons. Settings outside of
the indicated temperature range of 33 to 300 °C are not specifi ed and may cause
overload of the device.
The control time is about 30 minutes until 95% of the set temperature are reached. The
cooling time from +300°C to 100°C is approx. 40 minutes. The temperature calibrator
needs about 15 – 20 minutes until the specifi ed stability is reached once the set tempera-
ture is reached.
18

Advertenties

Inhoudsopgave
loading

Deze handleiding is ook geschikt voor:

40 92 13

Inhoudsopgave