14.2 Technical Description Concerning Electromagnetic Immunity
Table 2: Guidance & Declaration - electromagnetic immunity
The model CanalPro X-Move is intended for use in the electromagnetic environ-
ment specified below. The customer or the user of the model CanalPro X-Move
should assure that It is used in such an environment.
Immunity test
Electrostatic
discharge
(ESD)
lEC 61000-4-2
Electrical fast
transient/
burst
IEC 61000-4-4
Surge
lEC 61000-4-5
Voltage
dips, short
interruptions
and voltage
variations on
power supply
input lines
IEC 61000-4-
11
Power fre-
quency (50/60
Hz)
magnetic field
lEC 61000-4-8
NOTE: UT is the a.c. mains voltage prior to application of the test level.
32
IEC 60601
test level
±8kV contact
±2, ±4, ±8,
±15kV air
±2kV for power-
supply lines
±1kV for In-
put/output
lines
±0.5, ±1kV line
to line
±0.5, ±1, ±2kV
line to earth
<5 % UT
(>95% dip in
UT.)
for 0.5 cycle
<5 % UT
(>95% dip in
UT.)
for 1 cycle
70% UT
(30% dip in UT)
for 25 cycles
<5% UT
(>95 % dip in
UT)
for 250 cycles
30A/m
Compliance level
±8kV contact
±2, ±4, ±8, ±15kV
air
±2kV for power
supply lines
±0.5, ±1kV line to
line
±0.5, ±1, ±2kV
line to earth
<5 % UT
(>95% dip in UT.)
for 0.5 cycle
<5 % UT
(>95% dip in UT.)
for 1 cycle
70% UT
(30% dip in UT)
for 25 cycles
<5% UT
(>95 % dip in UT)
for 250 cycles
30A/m
Electromagnetic environ-
ment - guidance
Floors should be wood, con-
crete or ceramic tile. If floors
are covered with synthetic
material, the relative humid-
ity should be at least 30 %.
Mains power quality should
be that of a typical commer-
cial or hospital environment.
Mains power quality should
be that of a typical commer-
cial or hospital environment.
Mains power quality should
be that of a typical commer-
cial or hospital environment.
If the user of the models
CanalPro X-Move requires
continued operation during
power mains interruptions,
it is recommended that the
models CanalPro X-Move be
powered from an uninter-
ruptible power supply or a
battery.
Power frequency magnetic
fields should be at levels
characteristic of a typical lo-
cation in a typical commer-
cial or hospital environment.