Installation Qualification – IQ
ID

Description of installation

Ensure the cabinet is installed in-
I-1
doors.
Ensure the cabinet is installed in a
I-2
dry and sufficiently ventilated area.
Ensure the cabinet is not in direct
I-3
contact with sunlight or other heat
sources.
Ensure that the ambient operating
I-4
temperature is within the allowed
range.
Ensure that the cabinet is not
I-5
installed in a chloric/acidic
environment.
Ensure that the protective film on the
I-6
cabinet is removed.
Ensure that the cabinet is cleaned
I-7
with a mild soap solution
Ensure that the cabinet has stood
I-8
upright for 24 hours if the cabinet has
been laying down.
Ensure that the cabinet is levelled if
I-9
it is equipped with legs.
Ensure a level surface if the cabinet is
I-10
equipped with wheels/castors.
If equipped with wheels/castors:
I-11
Ensure wheels/castors are locked
after positioning
If equipped with drawers and/or
I-12
glass door: Ensure that tilt-bracket is
mounted.
Ensure a distance of 15-75 mm
I-13
between cabinet and back wall
Ensure that there is minimum a gap
I-14
of 30 mm between cabinets.
Ensure that the upper part of the
I-15
cabinet is not covered.
Ensure that electrical appliances are
I-16
not being used inside the cabinet.
Ensure connection from voltage-free
I-17
contact to external monitoring system
(optional).
58
Comply
Reference
in manual
YES
NO
page 8
page 8
page 8
page 8
page 8
page 8
page 8
page 8
page 9
page 9
page 9
page 10
page 10
page 11
page 11
page 12
page 13
Attachment