4.1.2.1.2.3 SOP Specific Conformance to Storage SOP Class
The IBox-DiS AE provides standard conformance to the Storage SOP Class. It shows the
following behavior on successful (with or without warnings) and unsuccessful transfer of
images:
Success (return status 0000)
Refused (return status A7xx) and error (return status A9xx or Cxxx)
Warning (return status B00x)
Messages indicating the status of a C-STORE request are displayed.
The transfer will run as a background process. Image acquisition is not delayed nor
interrupted.
Extended negotiation is not supported.
The standard conformance to the Storage SOP Class also implies standard conformance
for the
Ultrasound (US) Image Storage SOP Class
Ultrasound Multi-frame (US MF) Image Storage SOP Class
Table 4 lists the applied common Conditional (DICOM Type 1C and 2C) and Optional
(DICOM Type 3) attributes of these three Storage SOP Classes. All of these attributes
may be empty or not even present in certain situations.
DICOM specified usage: M = mandatory, U = user related
Informati
Module
on
Entity
Patient
Patient
Study
General
Study
Patient
Study
Series
General
Series
Equipment
General
Equipment
Image
General
Image
Image Pixel
SOP
Common
Cine
Multi-frame
Table 4: Applied Conditional and Optional Attributes of the US and US MF IOD
4.1.2.1.2.4 Association Acceptance Policy
The IBox-DiS AE does not accept associations.
Usage
US
US
MF
M
M
--
M
M
--
U
U
--
M
M
--
M
M
--
M
M
Content Date
Content Time
M
M
Planar Configuration
M
M
--
M
--
Frame Time
M
--
--
Conditional
Optional Attributes
Attributes
Patient Comments
--
Patient's Age
Patient's Size
Patient's Weight
Performing Physicians'
Device Serial Number
Software Versions
Image Type
Lossy Image
--
--
--
--
4 AE specifications
Name
Manufacturer's Model
Name
Compression
7