Download Inhoudsopgave Inhoud Print deze pagina

Vapormed Volcano medic Gebruiksaanwijzing pagina 27

Inhoudsopgave

Advertenties

10. HETELUCHTGENER ATOR
Guidance and Manufacturer's Declaration – Electromagnetic Immunity
The VOLCANO MEDIC Vaporization System is intended for use in the electro-
magnetic environment specified below. The customer or the user of the VOL-
CANO MEDIC Vaporization System should assure that it is used in such an en-
vironment.
Immunity test
IEC 60601
test level
Electrostatic
± 6 kV contact
discharge (ESD)
discharge
IEC 61000-4-2
± 8 kV air
discharge
Electrical fast
± 2 kV for power
transient/burst
supply lines
IEC 61000-4-4
± 1 kV for input/
output lines
Surge
± 1 kV line(s) to
IEC 61000-4-5
line(s)
± 2 kV line(s) to
earth
Compliance
Electromagnetic
level
environment – guidance
± 6 kV contact
Floors should be wood,
discharge
concrete or ceramic tile.
If floors are covered with
± 8 kV air
synthetic material, the
discharge
relative humidity should
be at least 30 %.
± 2 kV for power
Mains power quality
supply lines
should be that of a
typical residential,
not applicable:
commercial or hospital
input/output lines
environment.
are non-existent
± 1 kV line(s) to
Mains power quality
line(s)
should be that of a
typical residential,
± 2 kV line(s) to
commercial or hospital
earth
environment.
52
10. HETELUCHTGENER ATOR
Immunity test
IEC 60601
test level
Voltage dips,
<5 % Uτ
short
(>95 % dip in Uτ)
interruptions and
for ½ cycle
voltage
variations on
40 % Uτ
power supply
(60 % dip in Uτ)
input lines
for 5 cycles
IEC 61000-4-11
70 % Uτ
(30 % dip in Uτ)
for 25 cycles
<5 % Uτ
(>95 % dip in Uτ)
for 5 s
Power frequency
3 A/m
(50/60 Hz)
magnetic field
IEC 61000-4-8
Note: Uτ is the a.c. mains voltage prior to application of the test level.
Compliance
Electromagnetic
level
environment – guidance
<5 % Uτ
Mains power quality
(>95 % dip in Uτ)
should be that of a typical
for ½ cycle
residential, commercial or
hospital environment. If
40 % Uτ
the user of the VOLCANO
(60 % dip in Uτ)
MEDIC Vaporization
for 5 cycles
System requires
continued operation
70 % Uτ
during power mains
(30 % dip in Uτ)
interruptions, it is
for 25 cycles
recommended that the
VOLCANO MEDIC
<5 % Uτ
Vaporization System be
(>95 % dip in Uτ)
powered from an
for 5 s
uninterruptible power
supply.
3 A/m
Power frequency
magnetic fields should be
at levels characteristic of
a typical location in a
typical residential,
commercial or hospital
environment.
53

Advertenties

Inhoudsopgave
loading

Inhoudsopgave