Immunity test
Surge
Voltage dips
Voltage Interruptions
Conducted disturbances
induced by RF fields
Radiated RF Electromag-
netic fields
Electrical fast transients
/ burst
RATED Power frequency
magnetic field
Electrostatic Discharge
(ESD)
I F U
Basic EMC standard
IEC 60601-1-2 Edition 4 Test levels and compliance
or test method
Professional healthcare facility
environment
IEC 61000-4-5
± 0.5 kV, ± 1 kV
IEC 61000-4-11
0% U
At 0°, 45°, 90°, 135°, 180°, 225°, 270° and 315°
0% U
and
70% U
Single phase: at 0°
IEC 61000-4-11
0% U
IEC 61000-4-6
6V in ISM bands between
0.15 MHz and 80 MHz
80% AM at 1 kHz
IEC 61000-4-3
3 V/m
80 MHz – 2.7 GHz
80% AM at 1 kHz
IEC 61000-4-4
± 2 kV
100 kHz repetition frequency
IEC 61000-4-8
30 A/m
50 Hz or 60Hz
IEC 61000-4-2
Contact: ± 8 kV
Air: ± 2 kV, ±4 kV, ± 8 kV, ± 15 kV
Number of discharges: 10 for each polarity
; 0.5 cycle
T
; 1 cycle
T
; 25/30 cycles.
T
; 250 / 300 cycle
T
Dwell time: 3 sec
Frequency step size: 1%
Dwell time: 3 sec
Frequency step size: 1%
Home healthcare environment
6V in ISM and amateur bands
between 0.15 MHz and 80
MHz
80% AM at 1 kHz
10 V/m
80 MHz – 2.7 GHz
80% AM at 1 kHz
17