5. Technical Specifications
Control box
Batteries
Actuator
Hand Control
Weight
IP class
Material
Castors
Dimensions
Expected Service life
Max patient weight
Operating environment
Storage and transportation
environment
Materials in contact with
human skin
Applied parts:
P - patient,
C - caregiver
x
X
12
Linak, voltage 100V-240V, Max 1.5A, 50-60 Hz, IPX6, Class II (Double insulated)
Linak, 25,2V/2,1Ah, sealed,Li-Ion, 300mA IPX6, Cable length 200mm
Linak, DC 24 V, 4,3A, IPX6. Duty cycle: Max 10% or 2 minutes continuous use
followed by 18 minutes not in use, Push: 1500N. Stroke 400mm +/-2 mm
Cable length coiled 600 mm, IPX6
STD- 24 kg
STD S- 23 kg
STD LOW- 15 kg
IPX5
Steel
125/100/75mm
For dimensions, see "Dimensions" page
Walkers: 8 years
Textile products: 2 years under normal use. Two wash cycles a week are considered
as normal use. However, the total lifetime is dependent on the use, maintenance and
storage of the product.
150 kg / 330 Ibs
5 °C to 40°C, relative humidity 20% to 80% non-condensing, and atmospheric
pressure 700 to 1060 hPa.
5 °C to 50 °C, relative humidity 20% to 80% non-condensing, and atmospheric.
Environment pressure 700 to 1060 hPa.
Materials that patient and operator/caregiver are exposed to during usage of the
product for prolonged time:
Meditap B-365 (synthetic leather cushions) P
Polyurethane (on PU cushions) P
Dryflex (handles) P
PA6 + 20% GF (brake lever) P
Hand control C
X
Armrest P
Handles P
Knee pad P
Foot rest support P
Foot rest support clamps C
Handles adjusting knobs C
Hand controller and its holder. C
Handles to adjust the armrest width. C
Hand control C
Charger C
Charger cable C
Footplate P
x
I F U