BioCompact II
Tolerances
REFRIGERATORS (RR)
210 (Solid door)
+/- 3K
210 (Glass door)
+/- 3K
310 (Solid door)
+/- 3K
310 (Glass door)
+/- 3K
410 (Solid door)
+/- 3K
410 (Glass door)
+/- 3K
610 (Solid door)
+/- 3K
610 (Glass door)
+/- 3K
FREEZERS (RF)
210 (Solid door)
**
310 (Solid door)
**
410 (Solid door)
**
610 (Solid door)
**
*):
The temperature span between the initial temperature and the end temperature in the
hold-over test P-13,14.
**):
Please contact your local distributor for current information.
***): 90° opening 1 minute
Note:
Refrigerators (RR): Ambient temperature +25 °C
Setpoint temperature
Name:
Conducted by:
Inspected/verified by:
Door
Permissable
opening
deviation
recovery
(minutes)
time
(minutes)
***
4
35 %
4
10 %
4
35 %
4
20 %
8
15 %
7
10 %
6
35 %
6
35 %
15
20 %
11
10 %
19
10 %
10
20 %
Freezers (RF):
+5 °C
Signature:
Model: ________________________
Pull-
Permissable
Hold-over
down
deviation
23
5 %
5 °C → 10 °C
28
5 %
5 °C → 10 °C
24
5 %
5 °C → 10 °C
28
5 %
5 °C → 10 °C
21
5 %
5 °C → 10 °C
22
5 %
5 °C → 10 °C
15
5 %
5 °C → 10 °C
**
5 %
5 °C → 10 °C
53
10 %
-20 °C → -10 ° C
62
5 %
-20 °C → -10 ° C
71
10 %
-20 °C → -10 ° C
40
10 %
-20 °C → -10 ° C
Ambient temperature +25 °C
Setpoint temperature -20 °C
b i o l i n e
Hold-
Permissable
range*
over
deviation
66
43
55
42
47
33
**
**
56
56
42
**
Approved
Date:
(Yes/No):
SN: ________________________
5 %
5 %
5 %
10 %
10 %
5 %
10 %
10 %
5 %
5 %
10 %
10 %
111