Installation Qualification – IQ
ID
Description of installation
I-1
Ensure the cabinet is installed indoors.
Ensure the cabinet is installed in a
I-2
dry and sufficiently ventilated area.
Ensure the cabinet is not in direct
I-3
contact with sunlight or other heat
sources.
Ensure that the ambient operating
I-4
temperature is within the allowed
range.
Ensure that the cabinet is not installed
I-5
in a chloric/acidic environment.
Ensure that the protective film on the
I-6
cabinet is removed.
Ensure that the cabinet is cleaned with
I-7
a mild soap solution
Ensure that the cabinet has stood
I-8
upright for 24 hours if the cabinet has
been laying down.
Ensure that the cabinet is levelled if
I-9
it is equipped with legs.
Ensure a level surface if the cabinet is
I-10
equipped with wheels/casters.
If equipped with wheels/castors:
I-11
Ensure wheels/casters are locked
after positioning
If equipped with drawers and/or
I-12
glass door: Ensure that tilt-bracket is
mounted.
Ensure a distance of 15-75 mm
I-13
between cabinet and back wall.
Ensure that there is minimum a gap of
I-14
30 mm between cabinets.
Ensure that the upper part of the
I-15
cabinet is not covered.
Ensure that electrical appliances are
I-16
not being used inside the cabinet.
Ensure connection from voltage-free
I-17
contact to external monitoring system
(optional).
94
Comply
Reference
in manual
Yes
No
page 10
page 10
page 10
page 11
page 10
page 10
page 10
page 10
page 12
page 12
page 12
page 16
page 18
page 18
page 19
page 19
page 20
Model: ________________________
Attachment
Notes
SN: ________________________