Download Print deze pagina

Gram BioCompact II RF210 Gebruiksaanwijzing pagina 111

Advertenties

Appendix
BioCompact II
Tolerances
REFRIGERATORS (RR)
210 (Solid door)
210 (Glass door)
310 (Solid door)
310 (Glass door)
+/- 3K
410 (Solid door)
410 (Glass door)
610 (Solid door)
610 (Glass door)
FREEZERS (RF)
210 (Solid door)
310 (Solid door)
**
410 (Solid door)
610 (Solid door)
*
The temperature span between the initial temperature and the end temperature
in the hold-over test P-13,14.
**
Please contact your local distributor for current information.
Note:
RR: Ambient temperature +25 °C
Setpoint temperature
RF: Ambient temperature +25 °C
Setpoint temperature -20 °C
Name:
Conducted by:
Inspected / verified by:
Door
Permissable
opening
deviation
recovery
time
(minutes)
4
35 %
4
10 %
4
35 %
4
20 %
8
15 %
7
10 %
6
35 %
6
35 %
15
20 %
11
10 %
19
10 %
10
20 %
+5 °C
Model: ________________________
Pull-
Permissable
down
deviation
(minutes)
23
5 %
28
5 %
24
5 %
28
5 %
21
5 %
22
5 %
15
5 %
**
5 %
53
10 %
62
5 %
71
10 %
40
10 %
Signature:
b i o l i n e
Hold-over
Hold-
Permissable
range*
over
66
43
55
42
5 °C → 10 °C
47
33
**
**
56
56
-20 °C → -10 ° C
42
**
Approved
Date:
(Yes / No):
SN: ________________________
deviation
5 %
5 %
5 %
10 %
10 %
5 %
10 %
10 %
5 %
5 %
10 %
10 %
111

Advertenties

loading

Deze handleiding is ook geschikt voor:

Biocompact ii rf310Biocompact ii rf410Biocompact ii rf610