Deviations from the specifications dictated in the PQ are to be reported in the deviation report.
The PQ is concluded if all criteria of acceptance are approved and the possible deviations are rectified or accepted.
Measurement – Temperature stabilization
ID
Description
P-7
The test is intended to provide substantiation for the temperature stability inside
the cabinet during normal operation.
The temperature inside the cabinet must be stabilized – where all the points in the
working space have reached and maintained the same temperature.
When the system is stable, document ordinary operation of the cabinet at the
setpoint temperature and ambient temperature specified in P-5.
Duration:
The measurements throughout the operation test, must be documented and
attached the PQ.
Attachment:
Notes:
P-8
Are the measurements inside the allowed tolerances specified in P-6 ?
Attachment:
Notes:
Name:
Conducted by:
Inspected / verified by:
104
Signature:
Model: ________________________
Accepted
Yes
Approved
Date:
(Yes / No):
SN: ________________________
No