Deviations from the specifications dictated in the PQ, are to be reported in the deviation report.
The PQ is concluded if all criteria of acceptance are approved and the possible deviations are rectified or accepted.
Measurement – Prerequisites
ID
Description
P-4
Measurements made during the PQ tests must be documented and
attached to the PQ.
Attachment:
Notes:
P-5
Specify setpoint temperature:
Specify the ambient temperature: °C
Attachment:
Notes:
P-6
Allowed tolerances –
Select the tolerance, according to the model being tested.
Find model-specific tolerances in appendix.
Tolerance: +/-
Attachment:
Notes:
Name:
Conducted by:
Inspected / verified by:
°C
K
Model: ________________________
Signature:
b i o l i n e
Accepted
Yes
Approved
Date:
(Yes / No):
SN: ________________________
No
103